4 Deaths Linked To Thoratec Heart System

Thoratec Corporation today issued an urgent safety advisory about a serious problem with a key component of the HeartMate II LVAS system. The company said 4 patients had died and 5 patients had a loss of consciousness or other symptoms of hypoperfusion. The episodes occurred when patients and caregivers “experienced difficulties with the process of changing from a primary system controller to their backup system controller.”

The company said that 8 of the 9 events “occurred in patients who were converted to the Pocket Controller after being originally trained on an older model, the EPC System Controller….

Click here to read the full post on Forbes.

 